Specification Sheet
Fast Heat Up for Shorter Cycle Times
Readily available steam is fed into the
coils that surround the vertical autoclave
chamber. After a cycle is completed,
the autoclave remains in stand-by
mode,requiring zero waiting time
between cycles. The optional generator
signicantly reduces heat up times and
waiting time between cycles. This also
results in a faster,more effective cycle and
prevents the formation of condensate.
Improved Drying by Vacuum
The post-vacuum drying phase, at the end of the sterilization cycle,
ensures improved drying of porous loads and hollow instruments
such as pipette tips. The low pressure in the autoclave chamber
caused by the vacuum reduces the boiling temperature therefore
causing the moisture to evaporate faster. The hot vapor is sucked
out of the autoclave chamber by vacuum and the materials dry.
Following the vacuum stage, dry air is introduced into the chamber
through an air lter.
Complete Drying by Optional Steam Generator
The optional integral steam generator for the vertical autoclave
models, provides readily available steam that is fed to the coil that
surrounds the chamber. Based on the combined operation of heat
provided by the steam generator and vacuum air removal, the
leftover moisture will quickly evaporate from the chamber. This
results in a faster, more effective cycle and prevents the formation
of condensate and guarantees that even the most difcult loads
such as textiles, porous loads, hollow instruments and pipette tips
will dry.
Optional Drying
5
Moisture Removal by Vacuum Pump
Steam is injected into Coils
Autoclave Chamber is Heated