PRINTER OUTPUT DESCRIPTION
Autoclave No:01 Number of the autoclave with respect to other units in the
same location
Load number: 0003 Load number. Useful to determine when to clean the
chamber. (upon reaching 255 this number is reset to 0)
Operator :___________ To be filled in manually by operator.
O.K. Cycle completed successfully
−−−−−−−−−−−−−−−−−−−
D78 055°C 00P The time, temperature and pressure during drying.
*
* Prints drying data every 3 minute.
*
D21 104°C 00P The time, temperature and pressure during drying.
-----------------------
E21 112°C 04P The time, temperature and pressure during exhaust.
-----------------------
S21 121°C 17P The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S14 121°C 18P The time, temperature and pressure during sterilization.
S13 121°C 18P The time, temperature and pressure during sterilization.
----------------------
H12 117°C 14P The time, temperature and pressure during heating.
H00 038°C 00P The time, temperature and pressure during heating.
*
* Prints heat up data every 4 minute.
*
H00 72°F 00P The time, temperature and pressure during heating.
----------------------
MN TEMP PRES
DRY: 60MIN Drying time for selected program.
TIME: 08MIN Sterilization time for selected program.
TEMP: 121° Sterilization temperature for selected program.
PROG: PKG Selected program: Unwrapped instruments
TIME: 12:57:23 Time sterilization cycle begun.
DATE: 12:05:04 Date sterilization cycle begun.
Version: T04EWP Number and version of the program
Legend
MN Time elapsed in minutes
H Heating stage
S Sterilization stage
E Exhaust stage
D Drying stage
P psi