User Manual
Page 46 of 53 Pages
9.9 Cleaning Table Top Autoclaves with Chamber Brite ™
CHAMBER BRITE ™ is a cleaning and descaling agent designed
specifically for the cleaning and removal of water deposits, oxides and
other sediments that are found in steam sterilizers. The material is a
combination of acidic salts and additional cleaning materials. Chamber
Brite ™ autoclave cleaner has been formulated specifically to be a fast,
powerful and easy to use cleaner for steam sterilizers.
If the autoclave is not cleaned regularly dirt and debris will build up
and clog the tubing and solenoid valves. This dirt can also be
transmitted to the instruments during sterilization. In addition a layer
of dirt on the stainless steel chamber traps moisture against the metal
and will lead to the chamber becoming porous and failing.
It is recommended that your autoclave be cleaned with
CHAMBER BRITE ™ once per week
Caution!
NEVER
use bleach, steel wool, a steel brush or anything abrasive to
scrub or clean the Chamber
Cleaning Procedure
1. Important – all steps in this procedure
must be completed without
interruption.
2. When the autoclave chamber is cold,
remove instruments and trays from the
autoclave.
3. Open the door and spread the contents
of a packet in a straight even line along
the bottom of the chamber, from back
to front.
4 Select and start program No. 1.
(without dry
) When the cycle is
finished it will automatically exhaust
5. At the end of the exhaust cycle drain
the water from the reservoir.
6. Fill the water reservoir with distilled
water.
7. Repeat a sterilization cycle without
Chamber Brite ™ powder, to remove
any excessive dirt in the pipes. Select
and start program No. 1. (without dry
)
When the cycle is finished it will
automatically exhaust
8. At the end of the exhaust cycle drain
the water from the reservoir.
9. Turn the autoclave off and allow chamber to cool.