10. Remove the tray holder; rinse and wipe the interior of the chamber
with a damp cloth.
11. Fill the reservoir with distilled water or mineral free water only.
12. Press the manual water fill button and allow a small amount of
water (2-4 ounces) to fill chamber and flush out the fill tube.
Remove water from chamber.
13. The instrument is ready to use.
IMPORTANT:
DO NOT sterilize instruments during the cleaning process!!!
CAUTION:
Keep out of reach of children. Contains mildly acidic
ingredients. Avoid contact with the skin, eyes or clothing. Wash
hands well after touching the powder, in the case of eye contact
flush with continuous running water for at least 15 minutes. If
irritation persists get medical attention. If accidentally
swallowed, do not induce vomiting, drink large amounts of
water and obtain medical attention. MSDS available upon
request.
For models 1730, 2340, 2540 use one packet of CHAMBER BRITE ™.
For models 3545, 3850, 3870 use two packets of CHAMBER BRITE ™.
Clean every 20 cycles or as needed.