Manual

Set-up and Start-up
22 Hei-VAP Value, Hei-VAP Advantage
EN
EN
1. Remove the vapor tube from the packaging.
Figure 47: Install the vapor tube
2. Remove the threaded clamping screw (1) on the drive unit; remove the tension
spring (2) and PTFE 26 vacuum seal (3).
Figure 48: Locking button
3. Continue to press the locking button (Figure 48 (1)).
4. Remove the screw coupling connection of the vapor tube (Figure 4-7 (6) from the
drive unit.
5. Pull out the clamping sleeve (Figure 47 (5).
6. Push the vapor tube coupling clamp (Figure 47 (6)) and clamping sleeve (Figure 4-
7 (5)) with the bead first to the vapor tube (Figure 47 (7)). Verify that the bead au-
dibly snaps into the groove of the vapor tube (Figure 47 (7)).
7. Push the vapor tube (Figure 47 (7) into the drive unit with the pre-assembled parts.
8. Continue to press the locking button (Figure 48 (1)).
9. Manually tighten the vapor tube coupling clamp (Figure 47 (6)) with the threaded
connection of the motor drive (Figure 47 (4)) till a tight connection is made.
Locking buttons
3
4
5
6
7
1
2
1