Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control 103
Fig. 4-25: Installing the safety shield
1.
Hang the safety shield (1) on the edge of the heating bath (2) and position
it so that it does not interfere with the opening of the guard hood
(if present).
2. Turn eccentric screw (3) by 180° to clamp the safety shield to the heating
bath
Fig. 4-26: Boiling temperature sensor
The boiling temperature sensor can not be used when using the temperature
sensor T auto.
1.
Remove the screw cap on the condenser.
2. Insert the boiling temperature sensor in the condenser so that the white
PTFE side of the seal (2) faces the glass thread.
3. Secure the boiling temperature sensor with the screw connector cap.
4. Connect the bayonet plug (1) to the socket on the rear of the instrument
(see chapter
4.4.4, page 84).
Fig. 4-27: Temperature sensor Tauto
Only in conjunction with glassware sets G3, G4 and G6.
Safety shield
Boiling temperature
sensor (Laborota
4010 / 4011 and
4002 / 4003)
Temperature sensor
T auto (Laborota
4002 / 4003)