Instruction Manual

Table Of Contents
Using the Laborota 4002 / 4003 control
1.02
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control 115
6.4.5 Interrupting evacuation
To quickly interrupt the evacuation, e.g. in case of boiling delay or frothing:
Press Hold (
Fig. 6-1 (1)) button.
The vacuum valve closes or the vacuum pump stops (ROTAVAC vario
control). The LED next to Vac Start Stop (
Fig. 6-1 (13)) button flashes.
Press Hold (
Fig. 6-1 (1)) button.
The LED next to Vac Start Stop (
Fig. 6-1 (13)) button once again flashes.
6.4.6 Applying the current value as the setpoint
(p const mode only)
To apply the current value as the setpoint during evacuation:
Press Accept (
Fig. 6-1 (2)) button.
The current value is used as the setpoint for p const mode.
6.5 Time functions
6.5.1 Duration
The duration function records the time from switching on to switching off the
rotation.
1.
Press Select (
Fig. 6-1 (15)) until Set Time LED is lit.
2. Use the arrow buttons (
Fig. 6-1 (9)) to set the display (Fig. 6-1 (14)) to
zero.
Recording of the duration starts when rotation is started by pressing the
Rot Start Stop (
Fig. 6-1 (7)) or Auto Start Stop (Fig. 6-1 (3)) buttons.
Recording of the duration ends when rotation is stopped by pressing the
Rot Start Stop (
Fig. 6-1 (7)) or Auto Start Stop (Fig. 6-1 (3)) buttons.
The recorded time can be displayed by selecting Set Time within 30 seconds
after the rotation stops.
To display the duration during distillation:
1.
Press Select (
Fig. 6-1 (15)) until Set Time LED is lit.
The duration of distillation appears in the display (
Fig. 6-1 (14)) for five
seconds.
Reactivating the
vacuum
Setting the duration
function