Instruction manual ROTACOOL
E 2
E Key Page 2 1 2 3 4 Cover Cooling liquid connection Right threaded socket Left threaded socket Page 3 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Cup Display ‘Cooling’ LED ‘Pump’ LED On/Off button Main switch ‘SET’ button ‘’ button (decrease value) ‘’ button (increase value) Filling level indicator Extension kit Base of the housing Platform Sealing screw Discharge socket Discharge valve Unit socket Rotary evaporator Vacuum pump Connecting tube 3
E Thank you for buying this unit. You have acquired a product that was manufactured and tested by Heidolph in accordance with the highest quality standards. We hope it will enable you to carry out your work easily and to your complete satisfaction. CONTENTS PACKING LIST ..................................................................................................................... 4 OUTLINE DESCRIPTION ....................................................................................................
E OUTLINE DESCRIPTION CFC-free circulating chiller in a L-shaped design for Heidolph rotary evaporators. A Heidolph rotary evaporator (22) and a Heidolph vacuum pump (23) can be fitted on the circulating chiller to save space (Heidolph Instruments or your local Heidolph Instruments dealer will be glad to provide more information about suitable models). The circulating chiller is equipped with an air-cooled refrigerating machine and a circulating pump.
E SAFETY INSTRUCTIONS Follow the applicable safety regulations for your laboratory! Be extremely careful when dealing with inflammable substances. Before you connect the unit to your power supply, make sure that the supply voltage is the same as that on the rating plate. Make sure the unit stands securely! The unit is not explosion-proof. Caution must therefore be exercised when using it near flammable and explosive substances. ASSEMBLY 1. Important Read the following instructions before switching on.
E For temperatures under +10 °C, anti-freeze must be added to the coolant water. Check the highest and lowest permissible temperature of the selected thermal fluid against your desired operating temperature range. For thermal fluids, please note the following: Maximum operating temperature: The flash point of the thermofluid used should be at least 5 K higher. Minimum operating temperature: The viscosity of the thermofluid increases at low temperatures. This impairs flow and reduces heat transfer.
E 7. Filling the cooling liquid Only an externally closed system can be connected to the rotary chiller. Shut off the discharge valve (20) (so that the slit is vertical). Fill the circulation cooler via the filler neck after removing the cover (1) up to approx. 1 cm below the max. mark on the filling level indicator (14). The rotary evaporator cooler is vented by turning on the cooling and pump on the ROTACOOL with the On/Off button (9).
E 9. Attaching the extension kit Attach the extension kit (15) to the right hand side of the platform (17). 10. Setting up the rotary evaporator Place the rotary evaporator (22) securely on circulating chiller's platform (17). 11. Setting up the vacuum pump Place the appropriate vacuum pump model (23) on the circulating chiller. The vacuum pump is placed with its rubber feet in special cups (5) on the top of the circulating chiller in order to prevent accidental slipping.
E ROUTINE CLEANING AND SERVICING The cooling air for the refrigerating machine is sucked in through the liquefier in the base of the housing (16). The liquefier should be cleaned from time to time. To do this, tip the discharged unit onto its back. The fins of the black liquefier can now be cleaned using a brush or vacuum cleaner. Never use pointed or sharp-edged tools for this purpose, however, make sure the fins of the liquefier are not damaged or deformed as this will impair air flow.
E TROUBLESHOOTING Have any work required on the electrical, electronic and refrigerating components performed only by suitably qualified professionals. In the case of a malfunction, the unit will display a flashing error message on the screen. During error message output, the unit will switch between displaying the error message and the actual value of the flow temperature.
E TECHNICAL DATA ROTACOOL circulating chiller Operating temperature range Resolution/accuracy Temperature stability Cooling performance in W at an ambient temperature of 20°C [°C] [K] [K] Cooling liquid capacity Refrigerant in the chiller (provided by manufacturer) max. delivery (pressure) Delivery pressure (head) max.
E QUESTIONS / REPAIR WORK Questions Should you have further questions on the installation, operation or maintenance after reading the manual, please contact Heidolph Instruments or your local Heidolph Instruments Dealer. Repairs For repairs, please telephone Heidolph Instruments or your local Heidolph Instruments Dealer in the first instance.
Technischer Service – Technical Service – Service Technique – Servicio Técnico – Assistenza Tecnica DE Deutschland, Germany Heidolph Instruments GmbH & Co. KG Technischer Service Walpersdorfer Str. 12 91126 Schwabach Tel.: +49-09122-9920-74 Free Call: 0800-HEIDOLPH, 0800-588-9708 E-Mail: Stefan.Pill@Heidolph.de www.heidolph.de North America Phone: +1-866-650-9604 E-Mail: service@heidolph.com www.heidolphNA.com UK United Kingdom Phone: +44-01799-5133-20 E-Mail: sandra.mcneill@radleys.co.uk www.