6 Troubleshooting
Fault Possible cause Remedy
Pump does not start or
stops immediately.
Supply voltage too low or
powersupplyfailure?
Check supply voltage. Check
fuse.
Control signal for motor speed
ismissing?
Check control signal.
Pressure in outlet pipeline too
high?
Remove blockade in line,
open valve.
Motoroverloaded?
Allow motor to cool down,
identify and eliminate cause
of failure. Manual reset is
necessary. Switch off pump
or unplug mains.
Pump does not achieve
its ultimate vacuum or
usual pumping speed.
Leak in the pipeline or vacu-
umsystem?
Check pump directly - con-
nect vacuum gauge directly
at pump inlet - then check
connection, pipeline and
vacuum system if necessary.
Long,narrowline?
Use lines with larger dia-
meter, length as short as
possible.
Pumphasbeenexposedto
condensate?
Allow pump to run for some
minutes with atmospheric
pressure at the inlet.
Deposits have been formed
insidethepump?
Clean and inspect the pump
heads.
Diaphragms or valves dam-
aged?
Replace diaphragms and/or
valves.
Outgassing substances or va-
porgeneratedintheprocess?
Check process parameters.
Pump temperature too high
(motorspeedreduced)?
Ensuresufcientcoolingof
the pump or reduce inlet pres-
sure.
Pump too noisy.
Loudexhaustnoise?
Connect hose or silencer to
pump outlet.
Diaphragm crack or dia-
phragmclampingdiscloose?
Perform maintenance.
Other than above mentioned
causes?
Contact local distributor.
Pump seized.