User Manual

22 MMS
®
Inspection SPG High
Measurement
6 Measurement
Use the device model MMS Inspection SPG to measure the height of surface
profiles. The probe measures the peak-to-trough height differences according
to ASTM 4417, Method B. The gage is therefore suitable for height measure-
ments of surface profiles according to various directives and guidelines, such
as SSPC-PA17.
During the measurement you can also:
Delete measured readings: Press the key
View the statistics of the open measurement block: Press the BLOCK key
6.1 Notes on measuring
Essentially: If the specified precision is not achieved during the measure-
ment, you must recalibrate the gage, see page 24.
Measured readings outside the specified tolerance limits are displayed in
red and indicated by an illuminated red LED.
The gage vibrates during measurement acquisition if this is activated in the
gage settings.
6.2 Before you start
The gage is switched on ( key)
The required batch is opened ( key in the main menu, see also
page 21)
The probe tip has been inspected and shows no sign of damage. Examine
the probe tip for damage, see page 19.