User Manual

MMS
®
Inspection SPG High 25
Calibration
7.2.1 Calibration - procedure
1. Open the calibration: (in the measurement display)
2. Follow the routine and perform the calibration steps displayed (air value
and zero). In this regard, refer to the descriptions in the sections "Air
value calibration step", page 25 and "Zero calibration step", page 26.
3. Require a greater measurement accuracy?
Yes:
Continue the calibration routines with 1 or 2 calibration shims.
Observe "Foil calibration step (calibration shim)", page 28.
Calibration with only 1 calibration shim: Skip the calibration step
Foil 2 with .
No:
Skip the next two calibration steps (Foil 1 and 2) with .
Acknowledging (OK/ ) the last calibration step exits the calibration routine
automatically and the calibration is completed.
7.2.2 Air value calibration step
Measurement in air
Procedure air value calibration step
1. Hold the gage in the air and press OK.
The display automatically switches to the next calibration step or to
Measurement view (after switching on the gage or opening another
batch).