User Manual

28 MMS
®
Inspection SPG High
Calibration
7.2.4 Foil calibration step (calibration shim)
Measurements on the calibration shim (foil) located on the glass plate directly.
Required material
Calibration shim(s) from scope of supply. The hole in the shim marks the
specified measuring area.
Glass plate from scope of supply
Procedure Foil calibration step (calibration shim)
1 Position the calibration shim
Place the calibration shim (Foil 1/2) on the glass plate.
There must only be one calibration shim on the glass plate at any one time!
Glass plate
Calibration shim
Hole in the shim