User Manual

8 MMS
®
Inspection DFT High
Description
2.5 Gage concept
In order to measure, a Batch (file) and a suitable Calibration (reference) must
be created in the gage for the Measuring application. A description of the
terms measuring application, batch and calibration can be found in Chapter
"Glossary".
The essential content of the batch and calibration files and their relationship to
each other are shown schematically in the following illustration:
A measuring application is defined in the gage by a batch file and a calibration file
and their linking to each other.
Measuring application
Batch
(settings)
Limit monitoring
Grouping of measured read-
ings
Offset value
Counted average
Measurement acquisition
(automatic after placement
the probe on the sample,
manual)
Measurement view, decimal
places for the measured
reading
Unit of measurement
Save readings (yes/no)
Calibration
(recorded by reference measure-
ment)
Material properties of the
sample
Geometric shape of the sam-
ple
Surface roughness
1 or 2 reference coating
thicknesses
Air reference value
Gage