User Manual

20 MMS
®
Inspection DFT High
Settings for measurement
5.1 Creating a new batch
A Batch is created by selecting a Batch type. Each batch type is assigned a
certain Calibration method and contains presets for the measurement process.
Calibration and setup routines simplify creation of a batch.
Selection of the batch type and the associated calibration method is based on
the requirements of the guideline/standard or the accuracy of measurement
needed. An overview and description of the available batch types can be
found on page 11.
Before you start
The gage is switched on ( key)The appropriate batch template pack-
age has been selected in the gage settings for the primary field of use of
the gage. For a description, see page 11, in the Enhanced gage vari-
ants, only available with the corrosion batch template package upgrade.
Select batch template package: Main Menu () > Gage Settings > OK >
More Settings > OK > Usage
Reference part and/or calibration foils are prepared if you wish also to
create a new calibration with the new batch. Which parts you need de-
pends on the calibration method for the selected batch type, see
pages 11, 25 and 26.
Creating a new batch
1. Main Menu () > Batches > New
2. Select the desired batch type from the list
3. Tap on OK, to confirm your selection
4. Select an existing calibration from the list or New:
Only existing calibrations whose calibration method matches the selected
batch type are shown in the list.
5. Tap on OK, to confirm your selection
6. Follow the calibration routine
7. Continue with the setup routine and, depending on the selected batch
type, specify the rest of the measurement procedure (e.g. setting of limits,
block size). Confirm each setting with OK. After the setup routine has
been completed, the gage switches automatically to the measurement
view for the newly created batch.
Batch type Individual (IDV): Specify the rest of the measurement proce-
dure to suit your needs (e.g. setting of limits, block size). Confirm each