User Manual

22 MMS
®
Inspection DFT High
Settings for measurement
5.3 Assigning a different calibration to the opened batch
Before you start
The gage is switched on ( key)
The desired batch is opened, the corresponding batch name appears in
the colored header field
Assigning a new calibration
1. ‘Main Menu () > Batch Modify > OK > More Settings > OK > Calibration
Assign > OK
2. Select an existing calibration from the list:
Only the calibrations whose calibration method matches the open batch
are displayed.
3. Tap on OK, to confirm your selection
4. 2 x
> , switches to the measurement view for the opened batch
You have now assigned a different calibration to the opened batch.
What you can do next
Measure, see Page 23
Repeat calibration, recalibration, see page 29