User Manual

MMS
®
Inspection DFT High 29
Calibration
Acknowledging (OK) the last calibration step exits the calibration routine
automatically and the calibration is completed.
If measurements are to be carried out on both base materials (magnetiz-
able, non-magnetizable), repeat the calibration with the other base mate-
rial! Proceed as described in Chapter "Recalibration – Procedure".
What you can do next
Switch on the calibration function Quick, page 29
Assign the calibration to the opened batch, page 22
7.3.2 Recalibration Procedure
1. Open the calibration that is to be recalibrated: Main Menu () > Calibra-
tions > Name of calibration > CALIBRATE or (in the measurement data
display)
2. Follow the routine and perform the calibration steps displayed (Zero, Foil
1/2). In this regard, refer to the descriptions in the sections "Calibration
step Zero", page 30 and "Calibration step Foil", page 32.
Acknowledging (OK) the last calibration step exits the calibration routine
automatically and the calibration is completed.
If the Quick function (quick recalibration) is enabled, the calibration rou-
tine switches automatically to the next calibration step after 3 measure-
ments. A correction/change for the calibration foil used is not possible
when this function is enabled.
If measurements are to be carried out on both base materials (magnetiz-
able, non-magnetizable), repeat the calibration with the other base mate-
rial!
7.3.3 Fast recalibration
A semi-automatic calibration routine for calibration is activated with the Quick
function. If the Quick function is enabled, the calibration routine switches au-
tomatically to the next calibration step after every 3 measurements. Switching
to the next step is announced by a long signal tone. A correction/change for
the calibration foil used is not possible when this function is enabled.