User Manual

MMS
®
Inspection DFT High 31
Calibration
Procedure Calibration step Zero
1. Perform 5 to 10 measurements on the uncoated reference part.
NF/FE
calibration 003
0.0
x
OK
NF/FE
Calibration 003
OK
Zero
NF/FE
Calibration 003
OK
A Placing the gage
Place the gage on the un-
coated reference part.
B Acquiring the mea-
surement
An audible signal as well as
the lighting of the signal lamp
reports the measurement ac-
quisition. The new measured
reading appears in the dis-
play
C Lifting the gage
Raise the gage at least
25 mm above the surface
Repeat steps A to C for
the next measurement on the uncoated reference part
25 mm
-0.05
x
n
3
-0.01
x
Zero
-0.05
x
n
4
-0.01
x
Zero
-0.05
x
n
4
The display screens are to be understood only as examples
Uncoated
reference part