32 MMS
®
Inspection DFT High
Calibration
Display description - calibration step Zero
1 Measuring application (example)
2 Name of calibration (example)
3 Schematic illustration of the current
calibration step
4 Current calibration step
5 Progress display of the calibration
steps (example of 3 calibration steps,
calibration step 1 current)
6 Currently measured reading
(example)
7 Mean value of the existing number of
measured readings (example)
8 Delete last measured reading
9 Battery indicator
10 Number of measurements (example)
11 To next calibration step/exit calibration
7.3.5 Calibration step Foil
Measurement on the calibration foil lying directly on the uncoated reference
part.
Required material
Calibration foils with the desired thicknesses or from the scope of supply.
The circle on the foil marks the specified measurement area.
Reference part from customers own production, without the coating to be
measured
NF/FE
Calibration 003
Zero