User Manual

MMS
®
Inspection DFT High 33
Calibration
Procedure Foil calibration step (foil)
1. Placing the foil
Place the foil (1/2) on the uncoated reference part.
Only one foil at a time may lie on the uncoated reference part!
2. Perform 5 to 10 measurements on the calibration foil.
Uncoated
reference part
Calibration foil
Specified mea-
surement area
NF/FE
Calibration 003
OK
NF/FE
Calibration 003
OK
Uncoated
reference part
A Placing the gage
Place the probe of the gage
inside the circle on the foil.
B Acquiring the mea-
surement
An audible signal as well as
the lighting of the signal lamp
reports the measurement ac-
quisition. The new measured
reading appears in the dis-
play
C Lifting the gage
Raise the gage at least
25 mm above the surface
Repeat steps A to C for
the next measurement on the calibration foil
NF/FE
Calibration 003
OK
Calibration foil
25 mm
98.9
x
Foil 1
99.0
x
100
n
3
Foil
97.9
x
Foil 1
99.5
x
100
n
4
Foil
SET
SET
97.9
x
Foil 1
99.5
x
100
n
4
Foil
SET
The display screens are to be understood only as examples