User Manual

34 MMS
®
Inspection DFT High
Calibration
3. Enter the nominal value of the foil: SET > Use to set the nominal value of
the foil > OK
When using a calibration method with 2 foils, repeat the entire foil calibration
step (steps 1 to 4) with the second foil.
Display description - calibration step Foil
1 Measuring application (example)
2 Name of calibration (example)
3 Schematic illustration of the current
calibration step
4 Current calibration step
5 Progress display of the calibration
steps (example of 3 calibration steps,
calibration step 2 current)
6 Currently measured reading
(example)
7 Mean value of the existing number of
measured readings (example)
8 Delete last measured reading
9 Open window to enter the nominal value of the foil
10 Battery indicator
11 Nominal value of the foil entered under SET (example)
12 Quantity of measurements (example)
13 To next calibration step/exit calibration
4. Remove the foil from the reference part
Place the foil (1/2) back in the protective sleeve.
Uncoated
reference part
Calibration foil
NF/FE
Calibration 003