User Manual

MMS
®
Inspection DFT High 35
Calibration
7.4 Calibration- Assigning/changing names
Assign a unique name to the calibration (calibration method, material desig-
nation, batch no., …), example: Cal-1Foil EN AW 6082
Keep in mind that many calibrations with different calibration methods are
stored in the gage. A unique name makes it easier to select and assign
the desired calibration to a batch.
Procedure
1. Open the Rename function: Main Menu () > Calibrations > Name of cali-
bration > Rename
2. Change name:
Move cursor by means of the BACK and OK keys
Select the desired character by means of the keys
Delete character: Select space
3. Exit the Rename input window:
a Move the cursor entirely to the right using the OK key.
The character appears at the right edge of the display.
b Press OK to exit the input window.
7.5 Calibration - Reset
The reset function ensures that all coating thickness correction values for the
selected calibration are deleted. The Reset function deletes all parameters mea-
sured in the foil calibration step; the nominal values for the foil are retained,
however. Measurements in the zero calibration step are retained.
Procedure
1. Open the Reset function: Main Menu () > Calibrations > Name of cali-
bration > Reset > OK
2. Prompt asking whether you really wish to perform reset, confirm with YES.