6 Care, Cleaning and Maintenance
The Preamplifier needs no care and maintenance.
Cleaning
Do not use solvent to clean the housing.
7 Troubleshooting
Type of Malfunction Possible Causes Corrections
No LED lights. No 24 V supply.
Pressure sensor is defect.
Sensor cord is defect.
LED is defect.
Check connection cord, if
necessary change it.
Change pressure sensor.
Change sensor cord.
Loctite service.
Amplification is not adjustable
It lights only the same LED´s.
Potentiometer R10 is defect.
Loctite service
There is always an error message
at the controller, but the LED´s
light correct.
Preamplifier cord is defect.
Preamplifier is defect.
Change Preamplifier cord
Loctite service
There is always an error message
at the controller and no LED lights.
Amplification factor not correctly
adjusted.
Preamplifier cord is defect.
Pressure sensor is defect.
Sensor cord is defect.
Preamplifier is defect.
Adjust the amplification factor.
Change preamplifier cord.
Change pressure sensor.
Change sensor cord.
Loctite service.