11.6 Error codes
SSACK
Name
Description
Cause
Corrective action
NO
Normal
operation
Indicates the
success of the
requested
operation
no
EE
Execute
Error
Transponderdaten
und Lese-ID-
Sequenz können
nicht gelesen
werden
Reader processes previous
read or write request
Please wait until previous
query is finished
Transponder has no
or too few valid
ASCII characters in the MID
area
Program transponder with valid
ASCII characters in the MID area
Parameters for MID range
do not match the transpon-
der data
Adjust reader parameter for MID
area according to transponder
data area for MID
Wrong reader mode
(MANT / OP) for functional
operation
Switch to proper mode (MANT to
write MID)
CE
Communica
tion
Error
List format, list set
or data type is wrong
Check SECS message syntax
Syntax error with
message or
message
format or wrong
value
Data sent with a command
is incorrect
Check command syntax and data
Send parameter is not im-
plemented or out of range
Check parameter syntax and
value