User's Manual
UART Serial Port Data Output5.1
Table 3. Outputs Data Format
Serial Port Configuration
Data Format
Baud Rate: 57600; Data: 8 bytes; Stop: 1 byte; Parity: None.
After power-on, the detector will use around 3 seconds to complete the self-test. If there is an issue, the detector will
repeat self-testing until the problem is solved. After booting up, the serial port will send out data in the following format:
Header
$
Methane
Concentration
5 bytes
00318,
5 bytes
02147,
5 bytes
00735,
5 bytes
00999,
5 bytes
00036,
Light
Intensity
Environment
Temperature
Relevancy
*1000
5 bytes
00024,
5 bytes
-0078,
5 bytes
00000,
3 bytes
003,
3 bytes
Reserved
Bytes
SNR
000,
3 bytes
Version
Number
010,
3 bytes
Error
Code
000
Checksum
*3C
Data Description
Methane Concentration the unit is ppm*m
Light Intensity the light intensity of infrared detection laser
Relevancy reference value of whether the signal is distorted, <00980, no outputs
Environment Temperature measured from PCB board thermistor
Version Number used to record the software version number, e.g. 011
SNR signal to noise ratio reference value
UART Serial Port Data and Parameter Configuration
5