Owner's manual

CLX (1/12) Page 1
REV 5.1
Specifications
Measurement Range
0 – 10.00 mg/L (PPM)
Accuracy
±5% of reading or ±0.03 mg/L (PPM) whichever is greater for range of
0-6.0 mg/L(PPM)
±10% of reading from 6.01-10.00 mg/L (PPM)
Resolution
0.01 mg/L (PPM)
Cycle Time
Adjustable; 110 seconds to 10 minutes (600 seconds)
Note: the system defaults to 2.5 minutes
Display
Multi-Line Liquid Crystal Backlit Display
Alarms
Two Programmable, 120-240VAC 2A Form C Relay
Analog Output
Powered 4-20 mA, 600 drive, isolated
Communications Port
Bi-directional RS-485 with Modbus
Water Pressure
Integral pressure regulator 0.34 bar(5.0 PSI) to 10.3 bar (150 PSI.)
Flow Rate to Waste
200 – 400 ml/min.
Operating Temperature
0°C – 40°C (32°F – 104°F)
Wetted Materials
PVC, Borosilicate Glass, Reslyn (FFKM), Viton
®
(FKM),
Polypropylene, Stainless Steel, Acetal, Noryl
®
, Silicone
Sample Temperature Range
0°C – 55°C (32°F – 131°F)
Operating temperature for 30 day reagent life is 0-40°C (32°F – 104°F)
Reagent life will be <30 days if kept at temperatures above 40°C
Power Supply
100 – 240 VAC, 47 – 63 Hz, 150VA
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Environmental Conditions
Not recommended for outdoor use.
Altitude up to 2000 meters
Up to 95 % RH (non-condensing)
Enclosure Rating
Designed to meet IP 66 /NEMA 4X
Regulatory Compliance
And Certifications
CE Approved, ETL listed to UL 61010-1-2004 &
ETL Certified to CSA 22.2 No. 61010.1 2
nd
edition dated July 2004
Shipping Weight
3.9 kg (8.6 lbs.) Reagents Shipped Separately
Shipping Dimensions
406 mm X 406 mm X 241 mm (16”X 16” X 9 ½”)
Warranty
2 Years from date of manufacture