MICRO 200 BW TURBIDIMETER RANGE OPTION 0 - 1000 NTU HF scientific, inc. 3170 Metro Parkway Ft. Myers, FL 33916-7597 Phone: (239) 337-2116 Fax: (239) 332-7643 CATALOG NO. 21640 (5/05) Rev. 3.
Field Service Kit For the Micro 200 Series Turbidimeter Field Service Kit Contents: Qty. Description Field Service Kit, Catalog No. 19834 • Cat. No. 1 Sensor (Electronics Portion Only) 19807 1 Desiccant Tray Assembly 21555 1 Lamp Module 21541 1 Flow Head Assembly 50125 1 3 pk. cuvettes 50036 1 Secondary Standard Kit (0.02, 4, 10, 40, 100 NTU) 19820 1 O-ring Kit 50136 12 ft. Tygon Tubing 21201 Clean, Calibrate and Repair Your Instrument in the Field 1 Bulkhead Assembly, 2 pk.
FOREWORD HF TURBIDIMETERS HF turbidimeters are manufactured to meet design criteria for nephelometers as described in Standard Methods For Examination of Water and Wastewater. HF turbidimeters are approved by the U.S. EPA* as a means to measure the turbidity of potable water, waste water, and other liquids. HF turbidimeters provide a linear display of turbidity, throughout all ranges, in Nephelometric Turbidity Units (NTU).
DECLARATION OF CONFORMITY Application of Council Directive: 73/23/EEC Standard to which Conformity is Declared: Product Safety UL3111-1 CSA-C22.2 No.1010-1-92 CE EN61010-1 Immunity EN50082-1 IEC801-2 IEC801-3 IEC901-4 EMI EN55011 Group 1 Class A FCC Part 15 Class A Manufacture’s Name: HF scientific, inc. Manufacture’s Address: 3170 Metro Parkway, Fort Myers, Florida 33916-7597 Importer’s Name: Importer’s Address: Type of Equipment: Turbidimeter Model No.
DECLARATION OF CONFORMITY Application of Council Directive: 73/23/EEC Standard to which Conformity is Declared: Product Safety UL3111-1 CSA-C22.2 No.1010-1-92 CE EN61010-1 Manufacture’s Name: Immunity EN50082-1 IEC801-2 IEC801-3 IEC901-4 EMI EN55011 Group 1 Class A FCC Part 15 Class A HF scientific, inc. Manufacture’s Address: 3170 Metro Parkway, Fort Myers, Florida 33916-7597 Importer’s Name: Importer’s Address: Type of Equipment: Backwash Sensor and Interface Kit Model No.
MICRO 200 BW TABLE OF CONTENTS I. IMPORTANT INFORMATION ............................................................................................................................ 1 A. HOW TO USE THIS MANUAL ....................................................................................................................... 1 B. SPECIAL ENVIRONMENTAL CONSIDERATIONS ..................................................................................... 1 C. SPECIFICATIONS ..................................
MICRO 200 BW TABLE OF CONTENTS CONT. IV. ROUTINE OPERATION ........................................................................................................................................ 25 A. CONTINUOUS MONITORING FLOW THROUGH UNIT ............................................................................ 25 B. VAPORPURGE .................................................................................................................................................. 26 C. GRAB SAMPLES ..........
I. IMPORTANT INFORMATION A. HOW TO USE THIS MANUAL The equipment has been designed for simple and easy operation. In keeping with that philosophy, this MICRO 200 BW User’s Manual has been written to simplify all steps in the procedures that follow. THE MOST IMPORTANT ASPECT OF THIS MANUAL IS THAT YOU, THE USER, READ IT IN ITS ENTIRETY AND REFER TO IT OFTEN. Figures and drawings have been used throughout the manual to show you how and text has been used to enhance and clarify what you have been shown.
C. SPECIFICATIONS FOR THE MICRO 200 BW Specification Ranges: Method: Accuracy: Repeatability: Linearity: Resolution: 0 to 10 NTU 1 to 100 NTU Digital Display: Analyzer Graphics: Clock Graphics: Response Time: Averaging Time (Electronic): Microprocessor: Keyboard data entry system: Security Code: Built in diagnostics: Analog Output, Isolated: Alarms: Alarm Contact rating: Operating Temperature: Storage Temperature: Optical Cleaning for NTU Sensor: Flow rate: MICRO 200 BW 0 - 1000.
C. SPECIFICATIONS FOR THE MICRO 200 BW CONTINUED Specifications Shipping Weight: Warranty: Backwash Sensor - Optional: Backwash Sensor: Sensor operating depth: Standard cable length Sensor to Analyzer/Interface: Sensor to Analyzer distance: Backwash Sensor Dimensions: Backwash Sensor Case: Repeatability: Sensitivity: Linearity: Response time: Sensor operating temperature: Micro 200 BW Approximately 15 lbs. (6.8 kg) One year from date of shipment IR sensor measures 0 - 100% Transmittance 16 psi (32 ft.: 9.
E. FIGURES Figure No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Title Page No. Desiccant Tray Installation ......................................................................................... 6 Analyzer Outline Dimensions .................................................................................... 7 Turbidity Sensor Outline Dimensions ........................................................................ 7 Access Cover Removal .......................................
F. COMPLIANCE TESTING CERTIFICATION The Micro 200 BW has undergone extensive testing, therefore, HF scientific, inc certifies that this instrument has met the following standards. Product Safety US standard tested by ETL testing labs to UL 3111 Canadian Standard tested by ETL to CSA 1010.
II. INSTALLATION A. PACKING LIST FOR THE MICRO 200 BW CONTENTS QUANTITY must be mounted where it can be viewed at the backwash control station. In this example, for convenience of maintenance, the turbidity sensor should be mounted at the backwash station and have the water plumbed to it. The analyzer must be mounted in a location that avoids direct water contact. Refer to figure 2 on page 7 for outline dimensions.
Figure 2 Analyzer Outline Dimensions 8.062 XX.XX=INCHES (205) (XX)=mm Lamp LAMP Module HF HF## 21541 21541 MODULE 9.802 (249) 2.000 (51) 1.000 (25) 8.780 5.820 (225) (148) Figure 3 Turbidity Sensor Outline Dimensions MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
instrument. The desiccant pouch is designed to have a long life; however, replacement of the desiccant pack will be required from time to time. To determine the condition of the desiccant, check the included humidity indicator card. If required, a new foil sealed desiccant pouch and indicator card are available from HF scientific inc. part #21555R. It is essential that all enclosure seals be maintained.
the voltage. The correct fuse ratings are 5 x 20 mm 1A fast acting for 120 VAC and 5 x 20 mm ½A fast acting for 240 VAC. A power switch is located to the right of the power cord. WARNING: If power connection is to be hardwired, place each analyzer on a separate circuit breaker or switch to allow for service. Observe all local wiring codes. BW analyzer). 5. Recorder - Current A 4-20 mA current output is available. The connections are made at the terminal block labeled Analog 4-20 mA (J5).
WARNING: While the chance is slight, there are fault conditions under which a hazardous voltage could be exposed on the analog output wires when the jumper J13 is removed. To preclude this possibility HF scientific strongly recommends, for personal protection, that a Ground Fault Circuit Interrupter (GFCI), such as LEVITON part number 6599 or 5299, be used at the power connection.
before the alarm contacts can again close. One of the most important features of the MICRO 200 BW is the ease of operation. You simply press the Function Key, for the selection you want to access on the LCD screen. The information contained in this manual will help users take full advantage of the MICRO 200 BW’s capabilities. In the event that unusual circumstances arise that are not addressed in the manual, contact your local distributor.
2. Turbidity Display 9 6 1 5 8 7 3 ALARM F1 MENU F2 F3 F5 F4 2 4 This is the turbidity display. The display shown above is fairly typical. The display has the following parameters: 123456789- CURRENT TURBIDITY READING: 477.
3. Main Menu Main Menu F1 F4 ALARM Alarms F2 Display Parameters F3 Utility MENU Turbidity Backwash F5 Ver: M200BW-7B210ME6 This is the main menu. It is the starting point for all sub-menu selections. Shown below are all function key selections and their purpose. Selection Options: F1: Changes screen to ALARM MENU Select this option to set the value at which the alarm contacts will change for Alarm #1 & Alarm #2.
4. Alarm Menu Alarm Menu ALARM F1 F2 Alarm 1 1ØØØ. Off F3 Alarm 2 Ø2ØØ. Lo F4 Dead Band 1Ø% MENU F5 This is the ALARM MENU. The left side of the screen shows the selections as described below. The right side shows the current settings.
5. Display Parameters Display Parameters F1 Upper Limit 999. F2 Lower Limit ØØØ. F3 Select Time Base F4 ALARM MENU 6Ø Min. Set Average Select D to A 4 to 2Ø mA F5 This is the Display parameters Menu. The sub-menu selections are described below.
b. Set Average Set Average F1 ALARM X1 XØ1 F2 X16 F3 X48 F4 MENU X96 Digit Suppress Ø F5 This is the Set Average Menu. This is where you can make changes to the averaging period. Readings are updated at a rate of one per second. A higher averaging time will give smoother curves to the graphical display and to recording devices and helps “wash out” anomalies. Always use the X1 setting when taking grab sample readings.
a. Screen Brightness/History Screen Brightness/History ALARM F1 F2 Dimmer F3 Brighter F4 MENU F5 Clear History This is the screen brightness and clear history menu. In this menu you can adjust the screen intensity in increments by successively pressing F2 or F3. The intensity level will be stored when you leave this menu. Variations in ambient temperature will cause shifts in the brightness level. F4 allows you to clear the graph history from both the turbidity and backwash screens.
c. Calibrate 0 - 1000 NTU F1 Ø.Ø2NTU Calibrate Ø4Ø26 Hz (Current Frequency) ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ F2 1ØØ.ØNTU ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ 18Ø.6 F3 4ØØ.ØNTU ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ 4Ø2.5 ○ ○ ○ ○ ○ Ø.Ø227 ○ ALARM MENU 4Ø2.6 NTU (Current Reading) F4 1ØØØ.ØNTU ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ 1ØØØ.Ø F5 This is the turbidity calibration menu. Sensor Calibrate is provided to adjust the sensor output for variations in lamp intensity.
d. Comm Port Comm Port Menu F1 Set Baud F2 Set Address F3 Select Printer Printer Off F4 ALARM 9600 MENU 0 Access Code Menu F5 Reporting 4 Hour This is the communications or serial port menu. This menu will have no purpose unless you have one of the optional serial ports installed. See page 35 for more details.
f. Security Access Code Please enter Access Code ALARM F1 MENU F2 F3 F5 F4 This screen is encountered when leaving the turbidity or backwash displays and when changing the Access Code. Use only the F1, F2, F3, F4 or F5 buttons to enter the Access Code. Refer to page 30 for complete instructions. g. Set Access Code Set Access Code ALARM F1 F1 F2 F3 F2 Accept; press Menu===> MENU F3 F5 F4 This screen is encountered after the previous screen when changing the Access Code.
h. Self Test (4 - 20 mA Cal.) SELF TEST F1 F2 F3 GND +5 Ref. -5 Ref. 5V +15 +15 Sen. -15 -21 +24 .ØØØ1 5.ØØ6 -4.99Ø 4.978 14.99 15.Ø4 -14.79 -21.24 24.47 4 - 2Ø mA Cal. F4 ALARM MENU F5 This is the self test menu. Upon entering this menu all internal critical power supplies are checked and displayed. You may also calibrate the 4-20 mA D to A output while in this menu.
7. Wait Messages ALARM F1 F2 Please Wait MENU F3 Working . . . . . . . F5 F4 4 - 2Ø mA Cal. ALARM F1 MENU F2 F3 Working . . . . . . . F5 F4 Ø - 1ØV Cal. ALARM F1 MENU F2 F3 Working . . . . . . F5 F4 These screens indicate that the analyzer is performing some calculations or storing information. This typically occurs when the analyzer is powered up or when a change in display parameters has been made. MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
8. Backwash Display 9 6 5 1 8 7 3 OFF Ø9Ø.Ø Ø1Ø.Ø F1 ALARM MENU F2 F3 F5 F4 2 4 This is the backwash display. It is only meaningful if Backwash option is connected. The screen above shows a typical backwash cycle. In this mode, updates are made to the screen every 6 seconds. The reading is updated every second. The display shows the following parameters: 1CURRENT %T READING: 97.
a. Backwash Calibration Calibrate F1 ALARM Ø9984 Hz Ø99.8Ø% T F2 1ØØ.Ø%T ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ 1ØØØ5.Hz ○ MENU F3 F5 F4 This is the backwash calibration menu and usable only with Backwash Sensor connected. To get to it you must first enter backwash from the main menu. You will be asked if you wish to calibrate. If you respond with yes (F2) you will reach the calibration screen above. If you respond with no (F3) you will go directly to the backwash display menu.
B . START UP Before placing the MICRO 200 BW on line we recommend that you standardize the instrument. The following sections describe the required procedures. 2. Standardizing The reference standard (cuvette) supplied with the MICRO 200 BW is a pure liquid sealed in glass. It has a value of 0.02 NTU. 1. Indexing The United States Environmental Protection Agency (U.S. EPA) recommends that cuvettes used for instrument calibration or sample measurement be indexed.
The desiccant beads are dark blue when dry and turn to a light pink when expended. Check the desiccant on a quarterly basis or if after cleaning the cuvette readings appear erroneously high to ensure proper operation of the vaporpurge feature. It is important to keep a cover on the sample well if the flow through head is out for more than a few minutes, to avoid saturating the desiccant uselessly.
Figure 11 Vaporpurge MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
is accomplished by washing the interior and exterior of the cuvette in a detergent solution, then rinsing thoroughly 8 to 10 rinses in clean, distilled water, to remove all streaks. Flow shut-off clamp! " Flow Control Valve Figure 13 Lamp Replacement C. CALIBRATION PROCEDURES 1. Calibration Standards Figure 12 Critical Measurement Area a. Secondary Standard Set (optional) Catalog No.
b. Standard Formazin Solutions Calibration of this instrument is based on Formazin, a material which is made by polymerization. Calibration samples may be obtained by diluting Formazin stock suspension using “Turbidity-Free” water. Formazin stock suspension can be prepared by the user (Reference Standard Methods For Examination of Water and Wastewater) or a kit can be purchased from, HF scientific, inc., Catalog No. 50040.
VI. SECURITY ACCESS CODE The security access code for the MICRO 200 BW provides a degree of protection against an unauthorized change to the instrument’s settings or parameters. The security code feature may be turned off if desired. The first requirement for the security access code after “power-up” is when you wish to exit the current display screen (turbidity or backwash). Upon pushing the MENU button to leave the display screen, a request will be made for the security access code.
VII. TROUBLESHOOTING A. GENERAL NOTES Service and technical support are available from the manufacturer, HF scientific inc. from 8:00 AM to 4:00 PM Eastern Standard Time. The following Table will supply fundamental troubleshooting information. The Guide assumes that the controls and functions of the instrument are used correctly. The intention of the guide is to eliminate common faults, not to troubleshoot down to component level. GUIDE TO FUNDAMENTAL TROUBLESHOOTING SYMPTOM SOLUTION Display not lit.
B. SPARE PARTS LIST FOR MICRO 200 BW CATALOG NO.
VIII. OPTIONS chosen that best averages the backwash water over the cycle. The sensor should also be located as close to the backwash filter as possible to reduce the time delay of the water sample. This will give the operator a correct picture as to how the backwash is proceeding and when it is complete. Refer to figure 20 for sensor outline dimensions. BACKWASH A. INSTALLATION 1.
about three inches. Care must be used to not nick the wires under the jacket. Note that the sensor cable is composed of two pairs of wires. Each pair consists of two insulated wires and an uninsulated shield wire, wrapped individually in a foil shield. The wire pairs are RED & BLACK and WHITE & BLACK. The foil can be unwrapped and removed, while keeping the pairs together. The two shield wires should be wrapped together. The remaining wires need to have the insulation stripped back about ¼”.
c. Analog Outputs Analog outputs of either 4-20 mA or voltage are available. If you select voltage you may choose 0-10V, 0-1V or 0-100mV. This analog output will be shared by both modes and will automatically recalibrate to the selected mode. d. Serial Port (Optional) If a serial port is connected the readings will report in NTU while in the turbidity mode and will report in %T while in the backwash mode.
between the most remote MICRO 200 BW analyzer and the host computer does not exceed 50 feet (16.4 m.). Due to it’s “tri-state transmitter” design, the companion MICRO 200 BW units, which are on the same serial port at the host computer, may be “hubbed” or connected each one to the next as long as the total cable length from each MICRO 200 BW to the host computer does no exceed 50 feet (16.4 m.). See figure 23 on page 41. 4.
4-21/64 (110) 8-13/16 (224) 7-3/32 (180) 3-35/64 (90) Figure 16 Backwash Interface Outline Dimensions Figure 17 Backwash Interface Voltage Selection MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
4-21/64 (110) 8-13/64 (224) 7-3/32 (180) 3-35/64 (90) Figure 18 Junction Box Outline Figure 19 Junction Box Wiring Diagram MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
XXXX=INCHES (xx)=mm 7-21/32 (194) 5-7/16 (138) 5-1/8 5-21/32 (144) (130) 3.0 DIA. (76) Figure 20 Backwash Sensor Outline Dimensions SENSOR CABLE TO ANALYZER TAKES 1-1/2 IN. PVC PIPE SST HOSE CLAMPS SENSOR MOUNTING CHAMBER BRACKET SENSOR CABLE TO ANALYZER Figure 21 Backwash Channel Mounting MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
SENSOR CABLE TO ANALYZER PIPE(NOT SUPPLIED IN KIT) EXTENSION HANDLE MOUNTING KIT SENSOR INLET BAFFLE OUTLET BAFFLE OUTLET BAFFLE INLET BAFFLE TO CONTROL UNIT BACKWASH TROUGH FILTER BED Figure 22 Backwash Extension Handle Mounting MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
ADDRESS #1 ADDRESS #N RS232 INTERVENING RS232 VOLTAGE UNITS VOLTAGE 1ST OR HUB LAST OR STAR 1 4 5 S H I E L D HOST TXD CPU 1 8 GRD RXD 4 5 S H I E L D TXD 8 GRD RXD GRD TXD "BUS" OR "BACKBONE" Figure 23 RS-232 Voltage Connection ADDRESS #1 ADDRESS #N RS232 RS232 RS232 CURRENT CURRENT CURRENT INTERVENING "OUT" + "IN" - - "OUT" UNITS + + 1 2 3 4 5 6 7 8 "IN" - - HOST CPU S H I E L D + + + S H I E L D "IN" GRD GRD GRD - "OUT" + + "DAISYCHAIN" Figure 24 RS-
WARRANTY HF scientific, inc., as vendor, warrants to the original purchaser of the instruments to be free of defects in material and workmanship, in normal use and service, for a period of one year from date of delivery to the original purchaser. HF scientific, inc.’s, obligation under this warranty is limited to replacing, at its factory, the instrument or any part thereof.
GLOSSARY Alarms: Relay outputs that change when an alarm setpoint has been exceeded. JTU: Jackson Turbidity Units Backwash: A method of measuring the transmission of filter backwash water. LCD Display: Liquid Crystal Display (screen) used to view the information available on the MICRO 200 BW. Baud: Rate of transfer of serial information. Calibrate: Electronic adjustment of the instrument to known standards. Main Menu: The major options for each of the functions accessed by the “FUNCTION” keys.
MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.
Date: August 3, 2004 Revision: 10 MATERIAL SAFETY DATA SHEET HF scientific, inc. 3170 Metro Parkway Fort Myers, FL 33916-7597 (239) 337-2116 Mon-Fri 8:00 am - 4:30 pm 24 Hour Emergency Telephone Number CHEMTREC 1-800-424-9300 SECTION 1 PRODUCT IDENTIFICATION PRODUCT NAME: 100 %T STANDARD SOLUTION, 0.02 NTU PRIME TIME STANDARD, REFERENCE STANDARD, 0.02 NTU 250 ML PRIMARY STANDARD, and 0.02 NTU 1GALLON PRIMARY STANDARD, WATER - INCLUDING H2O PRIME TIME STANDARD. CATALOG NUMBER: HF scientific, inc.