Instructions

Temperature measurement range: 30-45
Working temperature: 10-35
Accuracy: ±0.5
Resolution:160 × 120
1
Installation-Tripod quick temporary deployment
Tripod Adapter For
Turret
A. Tripod connector fixed on the adapter
B. Fix the adapter on the camera
2
Configuration
C. Install the camera on the tripod
Installation Parameter Recommended
Model
Distance
(between human&camera)
DS-2TD1217B-3/PA
0.8-1.5m
DS-2TD2617B-3/PA
DS-2TD1217B-6/PA
1.5-3m
DS-2TD2617B-6/PA
3
Advice
Tripod Adapter for
Bullet
Turret:DS-2TD1217B
Bullet:DS-2TD2617B
1.Select Body Thermography as VCA Resource Type.
2. Go to Local Configuration interface, enable and save the
following settings:
3. Go to Body Thermography Basic Settings. Configure
and save the following settings:
Emissivity: For human skin, this value is normally set as
0.98.
Distance: The actual distance between the camera and
measured object.
Check Enable Face Detection and Display Temperature.
Suggest uncheck Upload Captured Face Image and Display Face
Temperature Position if its not a must.
Set Alarm Temp. and Pre-alarm Temp.
Press Draw Area to adjust face detection region.
Click Max. Pupil Distance and Min. Pupil Distance to draw
width filter frame, Max and Min distance refer to camera
detection distance.
4.Configure Camera 01 (Optical) Settings.
Uncheck Enable Blackbody Correction If no blackbody is used.
Check Enable of Body Temperature Compensation and keep
parameters as default.
Model Height
Elevation angle
DS-2TD1217B-3/PA
1.5m 20°
DS-2TD2617B-3/PA
DS-2TD1217B-6/PA
DS-2TD2617B-6/PA
D. Installation complete
The environment would easily influence the performance of thermal camera. Therefore, it is required to be
used in a stable indoor environment without wind, and the ambient temperature should be consistent.
Heres some advice:
1. Set up a one-way screening zone, and ensure that camera can clearly see the face of person whose body
temperature is being measured.
2. Avoid backgrounds that are too crowded or bright.
3. 30 minutes after turn on the camera, then go body temperature measurement, because the camera
itself needs to be steady first.
4. If there is a large gap between indoor and outdoor temperature, it is highly suggested to wait more
than 5 minutes then measure body temperature of whom just got in.
5. Follow the guidance of required temperature measurement distance.
Fever Screening Thermal Camera
Installation & Configuration Guide
5. Configure Camera 02 (Thermal) Settings.