Instructions

Fever Screening Thermal Camera
Installation & Configuration Guide
Temperature measurement range: 30-45
Working temperature: 10-35
Accuracy: ±0.5
Resolution:384 × 288
1
Installation-Wall-mounted deployment
2
Configuration
Installation Parameter Recommended
Model
Distance
(between
human&camera
)
Height Elevation angle
DS
-2TD2637B-10/P 2-7m
2.5m 20°
DS
-2TD2636B-15/P 2.5-9m
3
Advice
1.Select Body Thermography as VCA Resource Type.
2. Go to Local Configuration interface, enable and
save the following settings:
3. Go to Body Thermography Basic Settings. Configure
and save the following settings:
Emissivity: For human skin, this value is normally set as
0.98.
Distance: The actual distance between the camera and
measured object.
Check Enable Face Detection and Display Temperature.
Suggest uncheck Upload Captured Face Image and Display Face
Temperature Position if its not a must.
Set Alarm Temp. and Pre-alarm Temp.
Press Draw Area to adjust face detection region.
Click Max. Pupil Distance and Min. Pupil Distance to draw
width filter frame,Max and Min distance refer to camera
detection distance .
4.Configure Camera 01 (Optical) Settings.
Uncheck Enable Blackbody Correction If no blackbody is used.
Check Enable of Body Temperature Compensation and keep
parameters as default.
DS-2TD2636B/2637B
The environment would easily influence the performance of thermal camera. Therefore, it is required to be used
in a stable indoor environment without wind, and the ambient temperature should be consistent.
Heres some advice:
1. Set up a one-way screening zone, and ensure that camera can clearly see the face of person whose body
temperature is being measured.
2. Avoid backgrounds that are too crowded or bright.
3. 30 minutes after turn on the camera, then go body temperature measurement, because the camera itself
needs to be steady first.
4. If there is a large gap between indoor and outdoor temperature, it is suggested to wait more than 5 minutes
then measure body temperature of whom just got in.
5. Follow the guidance of required temperature measurement distance and height.
5. Configure Camera 02 (Thermal) Settings.

Summary of content (1 pages)