Service Manual

WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
29
1.1 Reader Replacement
14. Install the back cover on to the
reader. Make sure the gasket
inserts are correctly aligned.
15. Install the six screws to attach the back cover to the reader.
16. Remove any tape residue from the back cover of the reader.
17. Use the alcohol-based cleaner to clean the back cover of the reader
and the area where the reader was installed on the bed. Let the back
cover and area dry.
18. Remove one side of the adhesive covering from the double-sided
tape gasket, and install the gasket on to the back cover.
19. Remove the adhesive covering
from the other side of the
gasket, and install the reader on
the bed so that its label is
toward the outside of the bed.
Firmly press the reader into
position for 10 seconds with
approximately 75 lb (34 kg) of
force.
VersaCare® Bed shown
20. Connect the antenna-reader cables to the antenna cables.
21. Put insulated tape over the cable connections.
22. The replacement is complete. Do the function checks on page 53
and in the beds service manual. As you operate the bed’s
articulation functions, make sure the cables are not in a location
where they could get damaged or pinched.