Service Manual

WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
34
1.3 Indicator Light Assembly Replacement
1.3 Indicator Light Assembly Replacement
1. Make sure the bed is unplugged and you have done the steps for
“Setup” on page 25.
2. Raise the knee section to its highest position.
3. At the indicator light assembly, cut the light assembly cable to free it
from the light assembly.
4. Remove the indicator light assembly from the bed.
5. Use the alcohol-based cleaner to clean these areas, and then let the
areas dry:
The area where the indicator light was installed.
If the indicator light is being
replaced on a bed in the
state of California, the area
on the new indicator light
assembly where the Prop 65
label is to be installed.
The area on the back of the
new indicator light
assembly where a cable tie
mount is to be installed.
VersaCare® Bed—the mount is to be installed at the top
center of the assembly.
Progressa® Bed—the mount is to be installed in the center
of the assembly.
6. Install the Prop 65 label (if applicable) and cable tie mount on to the
indicator light assembly.
7. Remove the adhesive covering from the top of the new indicator
light, and install the indicator light so the label on the indicator light
is toward the outside of the bed. Firmly press the indicator light into
position for 10 seconds with approximately 75 lb (34 kg) of force.
8. Go to the section for your bed:
VersaCare® Bed” on page 35
“Progressa® Bed” on page 35
DRAFT 26-FEB-2019