WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
42
1.4 Antenna Assembly Replacement
1.4 Antenna Assembly Replacement
1. Make sure the bed is unplugged and you have done the steps for
“Setup” on page 25.
2. Make a note of these:
• The antenna’s location
• The antenna’s orientation
• The indicator color on the antenna’s cable.
NOTES:
• The antenna cable lengths are different. The indicator color
on the cable will help to make sure the new antenna is
correct for its location.
• The location and orientation of the antenna are very
important for the operation of the system.
VersaCare® Bed Progressa® Bed
3. Remove the antenna from the sleep deck, and set the antenna aside.
4. Use the alcohol-based cleaner to clean the area where the antenna
was removed. Let the area dry.