Service Manual

WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
51
1.6 Equipment Call Cable Replacement
WARNING:
Warning—Failure to correctly align the gaskets on the covers of the
reader could permit fluids to get into the enclosure. Injury or equipment
damage could occur.
15. Install the back cover on to the
reader. Make sure the gasket
inserts are correctly aligned.
16. Install the six screws to attach the back cover to the reader.
17. Remove any tape residue from the back cover of the reader.
18. Use the alcohol-based cleaner to clean the back cover of the reader
and the area where the reader was installed on the bed. Let the back
cover and area dry.
19. Remove one side of the adhesive covering from the double-sided
tape gasket, and install the gasket on to the back cover.
20. Remove the adhesive covering
from the other side of the gasket,
and install the reader on the bed
so that its label is toward the
outside of the bed.
Firmly press the reader into position for 10 seconds with
approximately 75 lb (34 kg) of force.
21. At the cable tie mount(s) near the reader, install a cable tie(s) to hold
the cables on to the bed.
22. Connect the WatchCare™ 1/4" communication cable to the bed and
to the equipment call jack.
23. The replacement is complete. Do the function checks on page 53 and
in the bed’s service manual. As you operate the bed’s articulation
functions, make sure the cables are not in a location where they
could get damaged or pinched.