WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
52
1.7 Label Replacement
1.7 Label Replacement
NOTE:
Replace a label if it can not be read or if it is not fully bonded to its
location.
1. Remove the label.
CAUTION:
Caution—Failure to clean the label location thoroughly could prevent the
label from bonding correctly.
2. Use the alcohol-based cleaner to clean the area where the label was
removed. Let the area dry.
3. Install the new label where the old label was removed. Make sure the
label has fully bonded to its location.
DRAFT 26-FEB-2019