Service Manual
Table Of Contents
- Table of Contents
- Intended Use
- Introduction
- Symbols
- Safety Instructions
- Quick View™ List of Features
- Prepare the System for Use
- Transport The Bed
- Cleaning and Disinfecting
- Preventive Maintenance
- Storage and Handling
- Expected Life
- Troubleshooting
- Replacement Parts
- Service Part Replacement Procedures
- Setup
- 1.1 Reader Replacement
- 1.2 Power Cable Replacement
- 1.3 Indicator Light Assembly Replacement
- 1.4 Antenna Assembly Replacement
- 1.5 Graphic Replacement for the Head, Seat, and Thigh Sections, and Deck Filler—Progressa® Bed
- 1.6 Equipment Call Cable Replacement
- 1.7 Label Replacement
- 1.8 WatchCare™ System Function Check
- Specifications
- Limited Warranty
WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
53
1.8 WatchCare™ System Function Check
1.8 WatchCare™ System Function Check
1. Make sure any WatchCare™ smart pads are at least 2 feet (61 cm)
from the bed, and then
plug the bed in.
2. At the indicator light, make sure the white light comes on solid. This
could take up to 2 minutes.
If the white light does not come on,
check the cable connections.
(For more information about the
indicator light, see “Indicator Light—Visual Identification” on page 8.)
3. Use the dry pad as follows to make sure the system operates correctly:
a. Install the sleep surface on the bed. Refer to the bed’s service
manual.
b. Put a dry pad on the bed in the correct location and orientation
(see “Prepare the System for Use” on page 9), and make sure you
hear a beep and the indicator light is solid green. This could take
up to 10 seconds.
NOTE:
If the indicator light does not operate as specified, see
“Troubleshooting” on page 18.