Service Manual

WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
58
Specifications
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV (100 kHz
repetition fre-
quency) for
Power Supply
Lines
± 2 kV (100 kHz
repetition fre-
quency) for
Power Supply
Lines
Mains power quality
should be that of a
typical hospital envi-
ronment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Ground
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Ground
Mains power quality
should be that of a
typical hospital envi-
ronment.
Voltage dips
IEC 61000-4-11
0% U
T:
0.5 cycle
At 0°, 45°, 90°,
135
°, 180°,
225
°, 270°, and
315
°
0% U
T:
1 cycle
and
70% U
T:
30
cycles
Single phase: at
0°
(See Note)
0% U
T:
0.5 cycle
At 0°, 45°, 90°,
135
°, 180°,
225
°, 270°, and
315
°
0% U
T:
1 cycle
and
70% U
T:
30
cycles
Single phase: at
0°
(See Note)
Mains power quality
should be of a typi-
cal hospital environ-
ment. If the user of
the P006979
requires continued
operation during
power mains inter-
ruption, it is recom-
mended that the
P006979 be pow-
ered from an unin-
terruptible power
supply.
Voltage inter-
ruptions
IEC 61000-4-11
0% U
T:
300
cycles
0% U
T:
300
cycles
Power Fre-
quency
Magnetic Fields
IEC 61000-4-8
30 A/m
60 Hz
30 A/m
60 Hz
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical hospital envi-
ronment.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P006979 is intended for use in the electromagnetic environment specified
below. The customer or the user of the P006979 should make sure it is used in
such an environment.
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment—
Guidance
DRAFT 26-FEB-2019