Service Manual

WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
59
Specifications
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P006979 is intended for use in the electromagnetic environment specified
below. The customer or the user of the P006979 should make sure it is used in
such an environment.
Immunity
Test
IEC
60601-1-
2 Test
Level
Compliance
Level
Electromagnetic Environment—
Guidance
Con-
ducted RF
IEC 61000-
4-6
3 V
0.15 MHz
- 80 MHz
6 V in
ISM
bands
between
0.15 MHz
and 80
MHz
80% AM
at
1 kHz
3 V
0.15 MHz -
80 MHz
6 V in ISM
bands
between
0.15 MHz
and 80 MHz
80% AM at
1 kHz
Radiated
RF IEC
61000-4-3
3 V/m
80 MHz
to 2.7
GHz
80% AM
at
1 kHz
10 V/m
80 MHz to
2.7 GHz
80% AM at
1 kHz
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey
a
, should be
less than the compliance level in
each frequency range.
Interference may occur in the vicin-
ity of equipment marked with this
symbol.
Note: These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and
people.
The P006979 is intended for use in the electromagnetic environment specified
below. The customer or the user of the P006979 should make sure it is used in
such an environment.
DRAFT 26-FEB-2019