User Guide

49
7.2 Troubleshooting
索引
4
3
2
1
7
6
5
10
9
8
Maintenance and Service
An incorrect ID number is
displayed
Is the test cable connected cor-
rectly?
When Open or No Shield is dis-
played because a cable is not
connected correctly, an incor-
rect ID number may be dis-
played. (Open or No Shield is
displayed when a wire is not
connected.)
Cable length is not
obtained when cable
length calibration has
been performed
Is the cable between 2 and 300 m
long?
The cable length measurement
range of this instrument is 2 to
300 m (6.6 to 984-ft.). However,
cable length cannot be mea-
sured if the cableÅfs characteris-
tic impedance is far from 100 Ω.
The mark is blinking
Replace the batteries (p.51).
Nothing is displayed even
when pressing the
POWER key
Replace the batteries (p.51)
Are the batteries installed cor-
rectly?
Check whether the batteries are
installed correctly.
If installed incorrectly, reinstall
the batteries (p.51).
Symptom Checks and Remedy
ON/OF
F