MEASUREMENT GUIDE 8807-01 8808-01 8807-51 8808-51 MEMORY HiCORDER
1 ______________________________________________________________________ Measurement Guide This Measurement Guide provides examples of settings for actual measurements using the 8807-01, 8808-01, 8807-51, 8808-51. The Guide provides only the minimum required information: please read the Operating Manual for usage in particular applications, and the notes about the measurement operations, before beginning to measure.
2 ______________________________________________________________________ Screen Configuration Environment SYSTEM Waveform display Comment DMM SYSTEM DISP DISP SET Initialize RS© SET SYSTEM DISP SET Settings on the screen MEM REC RMS STATUS STATUS STATUS CHANNEL CHANNEL CHANNEL SCALING SCALING SCALING TRIGGER (MEM) TRIGGER (REC) TRIGGER (RMS) Extended status Numerical calculation (MEM): Memory recorder function/ (REC): Recorder function/ (RMS): RMS recorder function ____
3 ______________________________________________________________________ (1) Measuring the Voltage Waveform of a 220 V AC Source (Memory Recorder Function) This example describes how to measure the voltage waveform of a 220 V AC (50 Hz) commercial power source using the Memory Recorder function. Move the flashing cursor using the cursor keys ( ), and use the and keys to make a setting. 1. 1. Press Press the the SET SET key key to to display display the the Status Status screen. screen.
4 ______________________________________________________________________ 3. screen. 1. Press Press the the SET SET key key to to display display the the Trigger Status screen. SYSTEM SET DISP Settings for Channel 1 (CH1) 1 1. Trigger mode: SING (single) (input the waveform once) 2 2. Trigger source: OR 3 4 3. Trigger type: LVL (Level) 5 4. Trigger level: 0.000 V 5. Slope: Trigger Setting The trigger activates at the rising zero crossing (0.000V) (Measurement is started.
5 ______________________________________________________________________ Waveform Display 1. Press the SET key to display the Status screen. When input is finished, the waveform appears on the Waveform Display screen. Waveform Print 1. Press the SET key to display the Status screen. PRINT COPY FEED Press the PRINT key to print out.
6 ______________________________________________________________________ (2) Measuring the RMS Voltage of a 220 V AC Source (RMS Recorder Function) This example describes how to measure the voltage waveform of a 220 V AC (50 Hz) commercial power source using the RMS Recorder function. Move the flashing cursor using the cursor keys and make a setting using the and keys. 1. 1. Press Press the the SET SET key key to to display display the the Status Status screen. screen. SYSTEM 1 2 SET DISP 1.
7 ______________________________________________________________________ 2. screen. 1. Press Press the the SET SET key key to to display display the the Channel Status screen. SYSTEM 1 2 3 SET DISP Settings for Channel 1 (CH1) 1. Waveform color : optional setting 2. Voltage axis range: 50 V (Measured by 50 Vx10 DIV) 3. Zero position: 0% (Set the bottom of the screen to 0 V) 3. screen. 1. Press Press the the SET SET key key to to display display the the Trigger Status screen.
8 ______________________________________________________________________ Waveform Display andtoPrint 1. Press the SET key display the Status screen. When measurement starts, the waveform is displayed and printed at the same time. NOTE When operating from the AC adapter, real-time printing requires that the time axis range be no faster than 1 s/DIV. When operating from the Battery Pack, real-time printing requires that the time axis range be no faster than 2 s/DIV.
9 ______________________________________________________________________ (3) Measuring a Current Waveform Using the 9018-10 (Memory Recorder Function) The 8807-01, 8808-01, 8807-51, 8808-51 MEMORY HiCORDER can measure only voltage by itself, but when connected with the 9018-10 CLAMP ON PROBE, it can also measure current. 1.1.Press Pressthe theSET SETkey keytotodisplay displaythe theStatus Statusscreen. screen. SYSTEM 1 SET DISP 1. Function: MEMORY 2 2. Time axis range: 5 ms 3.
10 ______________________________________________________________________ 3. screen. 1. Press Press the the SET SET key key to to display display the the Trigger Status screen. SYSTEM SET DISP Settings for Channel 2 (CH2) 1 1. Trigger mode: SING (single) (input the waveform once) 2 2. Trigger source: OR 3. Trigger type: LVL(Level) 4. Trigger level: 5.000 A 3 4 5 5. Slope: Trigger Setting The trigger activates at the rising 5.000A crossing. 5.000 A (Measurement is started.
11 ______________________________________________________________________ Waveform Display 1. Press the SET key to display the Status screen. When input is finished, the waveform appears on the Waveform Display screen.
12 ______________________________________________________________________ Waveform Print 1. Press the SET key to display the Status screen. PRINT COPY FEED Press the PRINT key to print out. Printing Example When the List&Gauge setting is set to "Gauge" or "List&Gauge" on the Environment screen, the gauge is printed with the waveform. The gauge is displayed as 'A' (Amperes) with the input coupling settings in effect.
HIOKI 8807-01, 8808-01, 8807-51, 8808-51 MEMORY HiCORDER Measurement Guide Publication date: October 2003 Revised edition 2 Edited and published by HIOKI E.E. CORPORATION Technical Support Section All inquiries to International Sales and Marketing Department 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL: +81-268-28-0562 / FAX: +81-268-28-0568 E-mail: os-com@hioki.co.jp URL http://www.hioki.co.