User Manual
1.2 Settings for Numerical Value Calculation
6
1.2 Settings for Numerical Value Calculation
1
2
Numerical Calculations
To open the screen: Press the SET key → Select with the SUB MENU keys →Num Calc Settings scree
n
Operating Key Procedure
1
Enable the Numerical Calculation function.
CURSOR
Move the cursor to the [Numerical Calc.] item.
F2
Select [On].
2
Specify the numerical calculation range.
CURSOR
Move the cursor to the [Calculation Area] item.
F1 to F8
Select either choice.
3
Perform calculation settings.
SHEET/PAGE
Select the group for which to set calculations.
(G1 to G16)
CURSOR
F1
Move the cursor to your setting choice, and se-
lect [Calculation].
4
Select the Calculation type.
CURSOR
Move the cursor to the [Type] column of the No.
row of the calculation to set.
F1 to F8
(Switch Display: F8)
Select the calculation type.
Whole Applies calculations to the whole
waveform. (default setting)
A-B Applies calculations to the data be
-
tween A/B cursors.
Calculation No.
Settings can also be made in
a dialog (
⇒ p. 8).
Calculation
Type
Group of Calculation
Settings
4
5
Calculation Setting
Choices
6
3
Calculation
Parameters
Setting choices depend o
n
the calculation type.
Channel for
Calculation
Off No calculation. (default setting)
Average Average value of waveform data
RMS Value RMS value of waveform data
P-P Value Peak-to-peak value of waveform data
Maximum Maximum value of waveform data
Time to Max Time from trigger to maximum value
Minimum Minimum value of waveform data
Time to Min Time from trigger to minimum value
Period Period of signal waveform
Frequency Frequency of signal waveform
When selecting [A-B], specify the calcula-
tion range using the A/B cursors on the
Waveform screen.
If no measurement data has been ac-
quired by the instrument, first measure
once so that the range can be specified for
calculations to be applied to subsequent
measurements.
Also select [Judgment] if you require judg
-