3.10 FFT Analysis Modes
101
Waveform Example
Power Spectrum Density (Linear Predictive Coding) LPC
When the spectrum shape is complex and hard to comprehend with either linear or power spectra, a
rough spectrum structure can be obtained.
Not available with external sampling enabled.
Main uses:
To obtain a spectral envelope using statistical methods
See About the Functions "3.10.2 Analysis Mode Functions" (⇒ p. 102)
Axis Display Type Description
X axis
Linear
Frequency is displayed with equal spacing
Display Range: DC to the top of the frequency range
Log
Frequency is displayed logarithmically
Display Range: 1/400
th
to 1/4000
th
(depending on the number of analysis
points) to the top of the frequency range
Y axis
Lin-Mag Analysis values are displayed linearly.
Log-Mag (logarithm) Analysis values are displayed as dB values. (0 dB reference value: 1eu
2
/Hz)*
* eu: engineering units that are currently set are the standard (e.g., when the unit settings is volts, 0 dB = 1 V
2
/Hz)
X axis: Log
Y axis: Lin-Mag
X axis: Log
Y axis: Log-Mag
• Always specify the order (from 2 to 64). Higher orders can expose finer spec-
tral details.
• Amplitude values provided by LPC are not always the same as the power
spectrum density.
• If an error occurs during analysis, no waveform is displayed.
• Noise-like phenomena can strongly affect the spectrum shape.