User Manual
1.2 Settings for Numerical Value Calculation
7
Operating Key Procedure
F1 to F8
(Switch Display: F8)
5
Select the channel for calculations.
CURSOR
Move the cursor to the [Ch] item.
F1 to F8
Select a channel for calculations.
The waveform calculations (Zn) can be selected.
6
Set parameters.
(not required for some calculation types)
CURSOR
Move the cursor to the [Parameter] item.
F1 to F8
Make appropriate parameter settings.
About setting choices (
⇒ p. 19)
See "3.3.3 Entering Text and Numbers" in the
Instruction Manual
7
Select a calculation group.
CURSOR
Move the cursor to the [Operand Selection] item.
F1 to F8
Select a calculation group.
Execute the calculations.
(when judging calculations (⇒ p. 12))
Applying Calculations to Existing Data
CURSOR
Move the cursor to the [Execute] button.
F1
Select [Execute].
When calculating automatically after measurement
START
Starts measurement.
*
Applicable to logic channels
Rise Time Rise time of waveform data
Fall Time Fall time of waveform data
Std Deviation Standard deviation of waveform data
Area Area enclosed by zero position and
signal waveform
X-Y Area Area of X-Y composite waveform
Time to Level* Time from trigger to specified level
Pulse Width* Pulse width of waveform data
Duty* Duty of waveform data
Pulse Count* Pulse count of waveform data
4 Operations Four arithmetic operations on numer-
ical calculation results
To print or save calculation results while
measuring
Before measuring, enable Auto Save (
⇒
p. 14) or Auto Print. Enable [Calc Results
]
on the Save Settings or Print Settings
screen.
See "10.3.4 Setting Auto Save",
"11.3 Making Auto Print Settings"
in the Instruction Manual
To print or save existing data
Press the PRINT or SAVE key (
⇒ p. 15).
Manual Print Settings
Manual Save Settings
See "11.4 Making Manual Print (PRINT
Key Output) Settings",
"10.3.5 Setting Manual Save (SAVE
Key Output)" in the Instruction
Manual
7
Execute calculation of the displayed
group.
Changes made to calculation settings
while measuring do not take effect unti
l
measurement has been stopped and re
-
started.
[Execute] button