User Manual
1.3 Judging Calculation Results
11
Set the judgment criteria (upper and lower threshold values) by which to judge
numerical calculation results. Judgment criteria can be set for every numerical
calculation.
Waveform acquisition processing depends on the trigger mode setting (Single or
Repeat) and the criteria specified to stop measuring upon judgment (GO, NG or
GO & NG ).
1.3 Judging Calculation Results
Make Calculation Settings
Start Measurement
Data Acquisition
Calculation
Display Calculation Results
Calculation Judgments
Print or Save
Stop Measurement
Stop Criteria
Trigger mode
Stop criteria not met
[Repeat]
Measurement repeats
until stop criteria are met
Measurement repeats until you
press the STOP key
[Single]
Stop Criteria Met or
Stop Criteria is GO&NG
Judgment when memory division is enabled
When memory division is enabled, waveform data is retained in the measured
block only when stop criteria are met.
When stop criteria are not met, measurement continues to repeat within the
same block.