3.5 Selecting Channels
71
Scaling _______________________________________________________
The scaling setting allows values displayed on this instrument to match the
actual values read directly on a sound level meter or vibration meter.
Setting example: To display measurement data on this instrument
so that it corresponds to that on a sound level meter.
In a case where a sound level meter displays 94 dB and the overall value
displayed on the Waveform screen of this instrument is -5.95 dB.
Without Scaling
With Scaling
Select [Ratio].
3
Move the cursor to [Ratio], and press the F7 [dB] key.
The dB Scaling dialog appears.
4
5
Enter the (overall) value “-5.95” dis
-
played on the instrument.
Enter the value "94" (from the soun
d
level meter) that you want to rea
d
directly.
6
Activate scaling.
Scaling is performed automatically, and the corresponding value
s
appear in the conversion ratio fields.
Sound Level Meter
Select [On].
2
1
See "Overall Value"
(
⇒ p. 109)
Overall: -5.95dB