2.2 Connecting the Cables
24
Before measuring voltage and connecting to the BNC jack, be sure to read "2.2.1 Connec-
tion Preparations" ( p. 16).
Connection Preparations
2.2.6 Connecting to the Model 8947 Charge Unit
Input module
Model 8947 Charge Unit
Maximum Input Voltage ( p. 4)
Measurement using a piezoelectric acceleration sensor:
Connect to the BNC jack or miniature receptacle
(according to the particular sensor).
Voltage measurement*
2
:
Connect to the BNC jack on an input module.
Model 9198 Connection Cord (recommended)
Preamplified acceleration sensor *
1
(BNC connector, 2-mA, 15-V power supply)
Charge-output type piezoelectric acceleration sensor *
1
(Miniature #10-32 connector)
*1. A commercially available acceleration
sensor can be used.
*2. For voltage measurement, the connec-
tion procedure is described in "2.2.2
Connecting to an Analog Unit (Models
8936, 8938, 8946, 8956 and 8957)"
(
p. 17).
Never connect the cable to the instrument while connected to a measure-
ment object. An electric shock could result.
The BNC jack for each channel and the miniature receptacles all share
common ground. To avoid short circuits, do not connect to two recepta-
cles at the same time.
To avoid electric shock or damage to the measurement object, turn the
instrument off while connecting a sensor or probe to the BNC jack.
When the [Preamp] measurement mode is selected, internal power (15 V
@ 2 mA) is applied to the BNC jack when measurement starts.
About preamplified acceleration sensors
The sensor should be compatible with the specification (15 V @ 2 mA) of the
Model 8947 Charge Unit. Incompatible sensors are likely to be damaged.