User Manual

2.6 Connecting Logic Probes
38
Maximum logic probe input voltages are as follows. Do not measure if
the maximum voltage would be exceeded, as damage the instrument or
personal injury may result.
Model 9327 Logic Probe: +50 VDC
Model 9320-01 Logic Probe: +50 VDC
Model 9321-01 Logic Probe: 250 Vrms (HIGH range), 150 Vrms (LOW
range)
Do not allow the metal tip of a logic probe to cause a short between con-
ductors on the measurement object. Never touch the metal tip of a probe.
Connection Procedure
Connect to the measurement object
See "Logic Probe Usage Procedures" ( p. 39)
Right Side
Example: Connecting the Model 9327 Logic Probe
1 Connect the logic probe by aligning the
groves on the plug and a LOGIC recepta-
cle.
2 Connect to the measurement object.
Required item:
Model 9327 Logic Probe
LOGIC receptacles
When a logic probe is not connected, the corresponding logic waveform appears
at HIGH level on the waveform screen.