User Manual

2.6 Connecting Logic Probes
40
For other models, refer to the specifications ( p. 84).
The number of channels 4 (Isolated between channels and the instrument)
Range LOW HIGH
Input resistance 30 k
Ω min 100 kΩ min
Sensitivity
Output: L 0 to 10 VAC
±(0 to 15) VDC
0 to 30 VAC
±(0 to 43) VDC
Output: H 60 to 150 VAC
±(20 to 150) VDC
170 to 250 VAC
±(70 to 250) VDC
Response time
Less than 1 ms Less than 1 ms
Less than 3 ms Less than 3 ms
with 100 VDC with 200 VDC
Maximum input voltage 150 V rms 250 V rms
Maximum rated voltage
to earth
250 V rms
Dielectric strength
(between unit and channels, between channels)
Insulation resistance
More than 100 MΩ / 500 VDC
(between unit and channels, between channels)
Using the Model 9321-01 Logic Probe
Connect to the measurement
object
Input selector
Connect to the LOGIC receptacle
Clips
1 Set the input selector in accordance with
the measured voltage.
LOW range: On/off for 100 VAC and 24
VDC, etc.
HIGH range: On/off for 200 VAC, etc.
2 Connect the clips to the measurement
object.
Inputs are non-polarized, so polarity can be disregarded.
Inputs are isolated, so each channel can be connected to a point of indepen-
dent potential.
This instrument detects absolute values, so DC of either polarity can be input.
AC voltage is evaluated as a 50/60 Hz standard sine wave.