User Manual

3.5 Model 8940 F/V Unit Settings
51
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(
p. 75).
[ 1:1, 10:1, 100:1, 1000:1(9322+9325) or 1000:1(9322+9418) ]
8
Set the threshold level
[Threshold]
Set the threshold value ( p. 72).
9
Set pull-up
[Pull-Up]
Enable (set On) the pull-up resistance when connecting to an
open-collector output signal. Disable it (Off) for normal mea-
surements (
p. 71).
[ Off or On ]
10
Set the Hold function
[Hold]
Select whether to hold the last measured value until the next
frequency or rotation rate is determined (
p. 70).
[ On, 10-ms Off or 1-s Off ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
Difference between frequency measurement and 50/60 Hz frequency
measurement
The frequency measurement function consists of selecting a range from 50 mHz
to 5 kHz and observing changes in frequency.
The 50/60 Hz frequency measurement function consists of setting either 50 or
60 Hz as the zero position, and observing deviations from that frequency.