3.8 Model 8958 16-Ch Scanner Unit Settings
60
3.8 Model 8958 16-Ch Scanner Unit Settings
Before Setting
The zero position of the Model 8958 16-Ch Scanner Unit needs to be adjusted in
the following cases.
See "Scanner Unit Zero Position Adjustment" (⇒ p. 27)
• When the Model 8958 16-Ch Scanner Unit has just been installed in the
instrument (a message requesting adjustment appears when the instrument is
turned on)
• When the zero position has shifted due to aging or environmental changes
3.8.1 Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (
⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Scanner].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (
⇒ p. 64).
Refer to the specifications for available ranges (
⇒ p. 99).
6
Set the digital filter
[Digital F]
Set the digital filter function to be enabled or disabled. Set it
to ON to suppress noise components (
⇒ p. 67).
[ Off, 60Hz, 50Hz or 10Hz ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
Execute to adjust inp
ut
signal offset. (⇒ p. 76
)
12 3
5
4
6