3.8 Model 8958 16-Ch Scanner Unit Settings
60
3.8 Model 8958 16-Ch Scanner Unit Settings
Before Setting
The zero position of the Model 8958 16-Ch Scanner Unit needs to be adjusted in
the following cases.
See "Scanner Unit Zero Position Adjustment" ( p. 27)
When the Model 8958 16-Ch Scanner Unit has just been installed in the
instrument (a message requesting adjustment appears when the instrument is
turned on)
When the zero position has shifted due to aging or environmental changes
3.8.1 Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item
Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (
p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Scanner].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (
p. 64).
Refer to the specifications for available ranges (
p. 99).
6
Set the digital filter
[Digital F]
Set the digital filter function to be enabled or disabled. Set it
to ON to suppress noise components (
p. 67).
[ Off, 60Hz, 50Hz or 10Hz ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
Execute to adjust inp
ut
signal offset. ( p. 76
)
12 3
5
4