User Manual

3.10 About Setting Contents
64
Select the appropriate measurement range. The setting value is the units per
division on the vertical axis.
Use the RANGE/POSN knob to make the setting independently from the cursor
position.
Setting Item: [Range]
•Refer to "Chapter 5 Specifications" ( p. 83) for the extents of the measure-
ment ranges.
If the range is undetermined, auto-range selection can be enabled.
Refer to "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the
Instruction Manual
When data has been stored in the instrument’s memory, an asterisk “
appears beside the data value’s measurement range.
When a portion of the input waveform is outside of the measurement range,
that portion is indicated by a different color on the display.
Voltage Axis Range
When variable auto adjustment is enabled [On] (default setting) and the variable
function is enabled, the variable setting becomes linked for measurement range
changes.
If you do not want the display area of the screen to change, disable variable auto
adjustment [Off].
See "12.2.7 Performing Variable Function Auto Adjustment" in the Instruction Manual
Temperature Measurement
About Measurement Range, Upper and Lower Input Limits and the Wave-
form Display:
The upper and lower limits of measurement input depend on the measurement
range.
Measurement is not available outside of the temperature range limits in the fol-
lowing tables:
3.10 About Setting Contents
3.10.1 Measurement Range Setting
Minimum Resolution
If the range is changed while measuring, acquired data is deleted and data
recording is restarted.
Model 8937 Voltage/Temp Unit
Measurement Range [°C/div] 10 20 50 100
Upper Limit [°C] 400 800 2000 4000
Lower Limit [°C] -92 -184 -460 -920
Model 8958 16-Ch Scanner Unit
Measurement Range [°C/div] 10 50 100
Upper Limit [°C] 200 1000 2000
Lower Limit [°C] -100 -200 -200