User Manual

3.10 About Setting Contents
69
This function periodically corrects for drift of the reference potential (about once
per second) to improve thermal characteristics in thermocouple mode.
A broken thermocouple wire can be detected during temperature measurement.
Normally when a thermocouple wire breaks, measured values exhibit random
instability.
3.10.8 Drift Correction Setting
Setting Item: [Drift]
Selections Description
Off
Drift correction is disabled. (Default setting)
On
Drift correction is enabled.
Data refresh rate is about once per second when drift correction is enabled.
Drift correction is available only while measuring. It does not affect the Level
Monitor on the Waveform and Channel Setting screens.
3.10.9 Disconnect (Burn-Out) Detection Setting
Setting Item: [Burn-Out]
Selections Description
Off
Broken wires are not detected.
On
Broken wires are detected.
When a broken wire is present, the waveform trace stays at the top
of the screen and the cursor value is displayed as “+OVER”.
Wire breakage is detected by sensing a miniscule current flow
(about 200 nA) through the thermocouple. If the thermocouple
wires are long or composed of a high-resistance material, set
[Burn-Out] to [Off] to avoid measurement errors.